Chapter 13 419
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:FREQ:STOP
Syntax :SENSe{[1]|2|3|4}:FREQuency:STOP <numeric>
:SENSe{[1]|2|3|4}:FREQuency:STOP?
Description Sets the stop value of the sweep range of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:FREQ:STOP 100E6"
20 OUTPUT 717;":SENS1:FREQ:STOP?"
30 ENTER 717;A
Related commands :SENS{1-4}:FREQ:STAR on page 418
Equivalent key
[Stop]
<numeric>
Description Stop value
Range 3E5 to 3E9
Preset value 3E9
Unit Hz (hertz)
Resolution 1