434 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SOUR{1-4}:POW:PORT:COUP
Syntax :SOURce{[1]|2|3|4}:POWer:PORT:COUPle {ON|OFF|1|0}
:SOURce{[1]|2|3|4}:POWer:PORT:COUPle?
Description Sets whether to output the same power level for each port of channel 1 (:SOUR1) to
channel 4 (:SOUR4). When the power slope feature is on (ON specified with the
:SOUR{1-4}:POW:SLOP:STAT command), the same power level is always outputted to
all ports regardless of this setting because different power levels cannot be outputted for
each port.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":SOUR1:POW:PORT:COUP OFF"
20 OUTPUT 717;":SOUR1:POW:PORT:COUP?"
30 ENTER 717;A
Related commands :SOUR{1-4}:POW:PORT{1-2} on page 435
Equivalent key
[Sweep Setup] - Power - Port Couple
Description
ON or 1 (preset value) Outputs the same power level to individual ports.
OFF or 0 Outputs different power levels to individual ports.