Chapter 13 371
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:AVER:COUN
Syntax :SENSe{[1]|2|3|4}:AVERage:COUNt <numeric>
:SENSe{[1]|2|3|4}:AVERage:COUNt?
Description Sets the averaging factor of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:AVER:COUN 4"
20 OUTPUT 717;":SENS1:AVER:COUN?"
30 ENTER 717;A
Related commands :SENS{1-4}:AVER on page 370
:SENS{1-4}:AVER:CLE on page 370
Equivalent key
[Avg] - Avg Factor
<numeric>
Description Averaging factor
Range 1 to 999
Preset value 16
Resolution 1