E5061A/E5062A commands
:SENS{1-4}:SWE:DEL
Syntax :SENSe{[1]|2|3|4}:SWEep:DELay <numeric>
:SENSe{[1]|2|3|4}:SWEep:DELay?
Description Sets the sweep delay time of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:SWE:DEL 0.05"
20 OUTPUT 717;":SENS1:SWE:DEL?"
30 ENTER 717;A
Equivalent key [Sweep Setup] - Sweep Delay
<numeric>
Description Sweep delay time
Range 0 to 1
Preset value 0
Unit s (second)
Resolution 0.001
Page view 423