416 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:FREQ:DATA?
Syntax :SENSe{[1]|2|3|4}:FREQuency:DATA?
Description For channel 1 (:SENS1) to channel 4 (:SENS4), reads out the frequencies of all
measurement point.
The data transfer format when this command is executed depends on the setting with the
:FORM:DATA command. (Query only)
Query response {numeric 1},…,{numeric NOP}<newline><^END>
Where NOP is the number of measurement points and n is an integer between 1 and NOP.
Example of use
10 DIM A(1:201)
20 OUTPUT 717;":SENS1:FREQ:DATA?"
30 ENTER 717;A(*)
Related commands :FORM:DATA on page 338
Equivalent key No equivalent key is available on the front panel.
Description
{numeric n} Frequency at the n-th measurement point