SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:MARK:FUNC:MULT:PPOL
Syntax :CALCulate{[1]|2|3|4}[:SELected]:MARKer:FUNCtion: MULTi: PPOLarity
{POSitive|NEGative|BOTH}
:CALCulate{[1]|2|3|4}[:SELected]:MARKer:FUNCtion: MULTi: PPOLarity?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), selects the polarity of the peak to be searched
with the multi peak search.
Parameters
Query response {POS|NEG|BOTH}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:MARK1:FUNC:MULT:PPOL NEG"
20 OUTPUT 717;":CALC1:MARK1:FUNC:MULT:PPOL?"
30 ENTER 717;A$
Related commands :CALC{1-4}:MARK:FUNC:MULT:PEXC on page 249
:CALC{1-4}:MARK:FUNC:MULT:TYPE on page 254
:CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key
[Marker Search] - Multi Peak - Peak Polarity
Description
POSitive (preset value) Specifies the positive peak.
NEGative Specifies the negative peak.
BOTH Specifies both the positive peak and the negative peak.