Agilent Technologies E5071C User's Guide Page 353

  • Download
  • Add to my manuals
  • Print
  • Page
    / 551
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 352
Chapter 13 353
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:MMEM:STOR
Syntax :MMEMory:STORe[:STATe] <string>
Description Saves the instrument state (data to be saved specified with the :MMEM:STOR:STYP
command) into a file.
Specify the file name with the .sta extension. If you want to specify a file on the floppy
disk drive, you need to add “A:” at the beginning of the file name. When you use directory
names and file name, separate them with “/” (slash) or “\” (backslash).
Notice that, if a file with the specified file name exists, its contents are overwritten. (No
query)
NOTE The instrument setting file saved with the autorec.sta (or A:autorec.sta) file name is
automatically recalled when turning on the E5061A/E5062A.
Parameters
Example of use
10 OUTPUT 717;":MMEM:STOR ""Test1/State01.sta"""
10 OUTPUT 717;":MMEM:STOR ""A:State01.sta"""
Related commands :MMEM:LOAD on page 345
:MMEM:STOR:STYP on page 363
Equivalent key
[Save/Recall] - Save State|Re-Save State
<string>
Description File name in which you want to save the instrument state
Range 254 characters or less