Chapter 5 83
Making a Measurement
Staring a Measurement Cycle (Triggering the instrument)
5. Making a Measurement
Staring a Measurement Cycle (Triggering the instrument)
Configuring the Instrument to Automatically Perform Continuous
Measurement
Step 1. Use the :INIT{1-4}:CONT command on page 341 to turn on continuous initiation mode
for the channels you want to measure and turn the mode off for any other channels.
Step 2. Issue the :TRIG:SOUR command on page 485 to set the trigger source to Internal trigger.
Starting Measurement on Demand
Step 1. Use the :INIT{1-4}:CONT command to turn on continuous initiation mode for the
channels you want to measure and turn the mode off for any other channels.
Step 2. Issue the :TRIG:SOUR command to set the trigger source to “Bus Trigger”.
Step 3. Trigger the instrument at any time you want to perform measurement. An external
controller can trigger the instrument using one of the following three commands:
Step 4. To start the next measurement cycle, repeat step 3.
Command
Can the *OPC? command on
page 198 command be used to
wait for the end of sweep?
Applicable
trigger source
*TRG on page 201
No
Bus trigger only
:TRIG on page 484 External trigger
Bus trigger
Manual trigger
:TRIG:SING on page 484 Yes