608 Chapter7
Language Reference
SRCPSWP Source Power Sweep
SRCPSWP Source Power Sweep
Syntax
Figure 7-255 SRCPSWP Syntax
Description
The SRCPSWP command activates and deactivates the power-sweep
function, where the output power of the tracking generator is swept
over the power-sweep range chosen. The starting source power level is
set using the SRCPWR command. The output power of the tracking
generator is swept according to the sweep rate of the spectrum
analyzer.
SRCPSWP is available only with an 8560E/EC Option 002.
Parameters
number real from 0 dB to 12.75 dB; 0.05 dB resolution via
GPIB.
UP or DN increments in steps equal to the value set by SRCPSTP.
Preset State
10 dB