Chapter 7 589
Language Reference
SAVES Save State
SAVES Save State
Syntax
Figure 7-234 SAVES Syntax
Description
The SAVES commandsavesthe currently displayed instrument state in
the specified state register.
Parameters
number integer from 0 to 9. Registers 8 and 9 are used by the
normalization routine. Numbers less than zero default
to zero; numbers greater than nine default to nine.
PWRON sets the spectrum analyzer to the current state when
the spectrum analyzer is turned on.
Example
10 OUTPUT 718;"IP;CF 300MHZ;SP 20MHZ;"
20 OUTPUT 718;"SAVES PWRON;"
30 END