378 Chapter7
Language Reference
ACPACCL Accelerate Adjacent Channel Power Measurement
ACPACCL Accelerate Adjacent Channel Power
Measurement
Syntax
Figure 7-4 ACPACCL Syntax
Description
The ACPACCL command sets the acceleration of the adjacent channel
power measurement to normal (NRM), faster (FASTR), or fastest
(FASTS). The ACP measurement techniques are changed when faster
and fastest are selected to speed up the measurement process. This
results in a degradation in the amplitude accuracy of 1 dB for faster
and as much as 2 dB for fastest. Refer to the softkey description of
ACCELERAT for more information.
Parameters
NRM, FASTR, FASTS
Preset State
NRM (normal)