Agilent Technologies 54503A User's Guide Page 416

  • Download
  • Add to my manuals
  • Print
  • Page
    / 716
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 415
416 Chapter7
Language Reference
AMPCOR Amplitude Correction
AMPCOR Amplitude Correction
Syntax
Figure 7-49 AMPCOR Syntax
Description
Use AMPCOR to turn the amplitude correction function on and off. The
ampcor function is used to compensate for frequency-dependent
amplitude variations. When ampcor is on, the current correction values
are added to all measurement results. Turning ampcor off does not
erase the current frequency-amplitude correction factors. Performing
an instrument preset (IP) or turning off the spectrum analyzer will turn
off ampcor.
Preset State
Off
Query Response
Figure 7-50 AMPCOR Query Response
Example
10 OUTPUT 718;"AMPCOR ON;"