Agilent Technologies 54503A User's Guide Page 301

  • Download
  • Add to my manuals
  • Print
  • Page
    / 716
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 300
Chapter 5 301
Programming
Program Timing
The next example processes trace information with a marker command,
MKPK HI (marker peak highest), which selects the highest amplitude
level in the trace. Because the program changes the measurement
range, the trace information must be updated with TS before MKPK HI
is executed. E
XAMPLE. Use TS to update trace information before using
the marker commands.
Connect the calibrator signal to the analyzer INPUT 50 before
performing this example.
10 CLEAR 718
20 OUTPUT 718;"IP;SNGLS;"
Initialize analyzer.
30 OUTPUT 718;"SP 1MHZ;CF 300MHZ;RL -5DBM;"
Change
measurement
range.
40 OUTPUT 718;"TS;"
Update trace.
50 OUTPUT 718;"MKPK HI;"
Place marker on highest signal
level.
60 END
Page view 300
1 ... 300 301 302 ... 716

Comments to this Manuals

No comments