Agilent Technologies 54503A User's Guide Page 241

  • Download
  • Add to my manuals
  • Print
  • Page
    / 716
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 240
Chapter 4 241
Key Function Descriptions
Key Descriptions
FULL SPAN Sets the spectrum analyzer to the center-frequency
span mode and sets the span to the maximum range.
Full span is:
Front-panel key access:
SPAN
GRAT ON OFF Blanks the graticule from the display (OFF) or
reactivates it (ON).
Front-panel key access: DISPLAY
GATE CTL EDGE LVL Selects the use of edge triggering or level
triggering to control the gate. In edge mode the gate
delay starts from the specified trigger edge. The gate is
then on for the time specified by the gate length
function.
In level mode the gate is controlled by the level of the
input trigger signal, so gate delay and gate length are
not functional.
Front-panel key access:
SWEEP
GATE DLY[ ] Controls the length of time from the trigger until the
gate is turned on. The second line of the softkey
indicates the current setting of the gate delay.
Front-panel key access:
SWEEP
GATE LEN[ ] Controls the length of time that the gate is on when
using edge triggering to control the gate. The second
line of the softkey indicates the current setting of the
gate length.
Front-panel key access:
SWEEP
GATE ON OFF Turns the gate function on and off. When set ON, the
analyzer activates the gate circuitry to function
according to the parameters controlled by gate length,
gate delay, and the gate trigger input.
Spectrum Analyzer Frequency Span
8560E/EC 2.9 GHz
Agilent 8561E/EC 6.5 GHz
Agilent 8562E/EC 13.2 GHz
Agilent 8563E/EC 26.5 GHz
Agilent 8564E/EC 40 GHz
Agilent 8565E/EC 50 GHz
Page view 240