270 Chapter4
Key Function Descriptions
Key Descriptions
RECALL AMPCOR Recalls a table of frequency-amplitude correction
points that was previously saved.
Front-panel key access: CAL
RECALL ERRORS Displays the last error that has occurred. Use the
step keys tocycle through accumulated errors. For alist
of all error codes and additional error information, refer
to Appendix C, "Error Messages", or to the installation
and verification manual.
Front-panel key access:
RECALL
NOTE When PRESET is pressed, the preselector data stored by the user does
not change. However, the factory settings now become active.
Factory preselector data always takes precedence over user-activated
preselected data, unless the user data is explicitly recalled using
RECALL PRSEL PK. For more information on storing and recalling
preselector data, refer to the SAVE PRSEL PK and RECALL PRSEL
PK softkeys.
RECALL OPN/SHRT Recalls the internally-stored open/short average
calibration trace into trace B. The instrument state is
also set to the state of the stored open/short calibration.
Front-panel key access:
AUX CTRL
RECALL PRSEL PK Only for use with internal mixing or for preselected
external mixing. Recalls the preselector data that is
stored by the user in the user data table. This data can
be saved using SAVE PRSEL PK, found under the
SAVE
menu. The three available tables of preselector data are
described below.
1. Current Data Table contains the data used to tune
the preselector. The data is obtained from either the
user data table or the factory data table. Use either
the PRESEL MAN ADJ or PRESEL AUTO PK
softkey (located under
AMPLITUDE or AUX CTRL) to
modify the data in the current table. To save the
data, use SAVE PRSEL PK. If the current data is
not saved, it is lost when the instrument is preset,
turned off, or whenever bands are changed in the
case of external mixing.
2. Factory Data Table is the default data table for
internal mixing. This data is set at the factory and
can be changed only by service personnel. This data
is copied to the current data table when
PRESET is
pressed.