Page view 520
Chapter 7 521
Language Reference
MKMCF Marker Mean to the Center Frequency
MKMCF Marker Mean to the Center
Frequency
Syntax
Figure 7-169 MKMCF Syntax
Description
The MKMCF command moves the midpoint of the two displayed
markers to the spectrum analyzer center frequency. This command is
useful when making occupied channel power measurements. Use the
MKDELCHBW command to change the desired channel power to the
power between the current delta markers. The MKMCF command can
then be used to center this bandwidth on the display for making further
measurements.
Example
10 OUTPUT 718;"MKMCF;"