Agilent Technologies 54503A User's Guide Page 380

  • Download
  • Add to my manuals
  • Print
  • Page
    / 716
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 379
380 Chapter7
Language Reference
ACPALPHA Adjacent Channel Power Alpha Weighting
ACPALPHA Adjacent Channel Power Alpha
Weighting
Syntax
Figure 7-6 ACPALPHA Syntax
Description
The ACPALPHA command is used to set the alpha weighting for an
adjacent channel power measurement.
Parameters
number unitless real number between 0 and 1
Query Response
Figure 7-7 ACPALPHA Query Response
Example
10 REAL Alphaweight
20 Alphaweight = 0.13
30 OUTPUT 718;"ACPALPHA";Alphaweight;";"
40 END