Chapter 2 153
Making Measurements
Example 11: Time-Gated Measurement
Displayed spectrum too low
in amplitude.
Resolution bandwidth or
video bandwidth filters not
charging fully.
Widen resolution
bandwidth or video
bandwidth, or both.
Displaychangeddrastically
during operation.
Gate functions have been
turned off by changing
detector path. For example,
activating the marker noise
function (MKRNOISE ON
OFF) changes the detector
path.
Do not change the detector
path during the
measurement.
Table 2-9 If You Have a Problem with the Time-Gated Measurement