Page view 702
Chapter 9 703
If You Have a Problem
Error Messages
Error codes 700 through 799 relate to digital and checksum
failures. Instrument service is required.
ERR 700 EEROM Checksum error of EEROM A2U501.
ERR 701 AMPL CAL Checksum error of frequency response
correction data.
ERR 702 ELAP TIM Checksum error of elapsed time data.
ERR 703 AMPL CAL Checksum error of frequency response
correction data.
ERR 704 PRESELCT Checksum error of customer preselector peak
data; external preselector data recalled in internal
mode; or internal preselector data recalled in external
mode.
ERR 705 ROM U306 Checksum error of program ROM A2U306.
ERR 706 ROM U307 Checksum error of program ROM A2U307.
ERR 707 ROM U308 Checksum error of program ROM A2U308.
ERR 708 ROM U309 Checksum error of program ROM A2U309.
ERR 709 ROM U310 Checksum error of program ROM A2U310.
ERR 710 ROM U311 Checksum error of program ROM A2U311.
ERR 711 RAM U303 Checksum error of system RAM A2U303.
ERR 712 RAM U302 Checksum error of system RAM A2U302.
ERR 713 RAM U301 Checksum error of system RAM A2U301.
ERR 714 RAM U300 Checksum error of system RAM A2U300.
ERR 715 RAM U305 Checksum error of system RAM A2U305.
ERR 716 RAM U304 Checksum error of system RAM A2U304.
ERR 717 BAD uP!! Microprocessor is not fully operational.
ERR 718 BATTERY? Nonvolatile RAM is not working; check
battery.
ERR 719 MODEL #? Cannot identify the model number.
ERR 720 OPTION #? Cannot identify an option number.
ERR 721 AMPC RAM Checksum error of the ampcor function
correction data.
ERR 750 SYSTEM Hardware/firmware interaction, zero divide;
check other errors.