SRCPWR Source Power
Parameters
number real from 10 dBm to +2.8 dBm; 0.05 dB resolution via
GPIB.
UP or DN increments in steps equal to the value set by SRCPSTP.
Preset State
10 dBm
Query Response
Figure 7-258 SRCPWR Query Response
Example
10 OUTPUT 718;"IP;SNGLS;"
20 OUTPUT 718;"FA 300KHZ;FB 1GHZ;"
30 OUTPUT 718;"SRCPWR ON;"
40 OUTPUT 718;"SWPCPL SR;"
50 PRINT "CONNECT THRU. PRESS CONTINUE WHEN READY TO STO
RE."
60 PAUSE
70 OUTPUT 718;"SRCTKPK;DONE?;"
80 ENTER 718;Done
90 OUTPUT 718;"TS;DONE?;"
100 ENTER 718;Done
110 OUTPUT 718;"STORETHRU;"
120 OUTPUT 718;"TS;DONE?;"
130 ENTER 718;Done
140 OUTPUT 718;"NORMLIZE ON;"
150 OUTPUT 718;"TS;DONE?;"
160 ENTER 718;Done
170 OUTPUT 718;"NRPOS 5;TS;"
180 PRINT "RECONNECT DUT. PRESS CONTINUE WHEN READY."
190 PAUSE
200 OUTPUT 718;"NRL -10DB;"
210 OUTPUT 718;"TS;DONE?;"
220 ENTER 718;Done
230 LOCAL 718
240 END