
424 Chapter7
Language Reference
AT Input Attenuation
AT Input Attenuation
Syntax
Figure 7-60 AT Syntax
Description
The AT command sets the amount of attenuation between the input
and the first mixer. The attenuation can be set to 0 dB only by numeric
data entry, and not by using the knob or step keys.
Parameters
number integer from 0 to 70, in decade increments. Numbers
are rounded up to the nearest decade. (integer from 0 to
60 for the Agilent 8564E/EC and Agilent 8565E/EC)
AUTO sets the attenuation to coupled mode.
MAN sets the attenuation to manual mode.
UP or DN changes the attenuation by 10 dB (but cannot set
attenuation to 0 dB).
Preset State
Coupled mode, 10 dB
Comments to this Manuals