Chapter 7 431
Language Reference
BML Trace B Minus Display Line
BML Trace B Minus Display Line
Syntax
Figure 7-67 BML Syntax
Description
The BML command subtracts the display line from trace B and places
the result in dBm (when in log mode) in trace B, which is then set to
view mode. In linear mode, the results are in volts.
NOTE The displayed amplitude of each trace element falls into one of 600 data
points. There are 10 additional points of overrange, which corresponds
to one-sixth of a division. When adding or subtracting trace data, any
results exceeding the limits are clipped at that limit.
Example
10 OUTPUT 718;"IP;"
20 OUTPUT 718;"BLANK TRA;"
30 OUTPUT 718;"CLRW TRB;TS;DL -50DBM;"
40 OUTPUT 718;"BML;"
50 END