688 Chapter9
If You Have a Problem
Error Messages
Error Messages
Error messages are displayed in the lower right-hand corner of the
analyzer display. These error codes are provided for service personnel
who troubleshoot the spectrum analyzer. However, they also alert the
user to errors in spectrum analyzer function or use. Error codes are set
up in these categories:
100 to 199 Programming errors
200 to 299 ADC failures
300 to 399 LO or RF failures
400 to 599 IF failures
600 to 699 Display failures
700 to 799 Digital failures
800 to 899 Option module failures
900 to 999 User generated errors
If returning the analyzer for repair is necessary, include a list of any
error codes and messages that appeared when the trouble began. These
codes and messages provide troubleshooting information for the service
person working on the spectrum analyzer.
Viewing Multiple Messages
The spectrum analyzer displays only one error message at a time. More
error messages may exist, since one problem can cause several errors.
Check for more error messages and find the one that is the most likely
cause.
1. Press
RECALL, then MORE 1 OF 2.
2. Press
RECALL ERRORS. An error message is displayed in the active
function block on the analyzer display.
3. Use the and keys to scroll through any other error messages
which might exist, making note of each error code.