5-11. Battery measurement
The internal resistance of a battery is generally measured using a 1 KHz AC signal. When a battery
is connected directly to the auto balancing bridge type impedance measurement instrument, the
instrument becomes the DC load, typically 100 for the battery. Figure 5-37 shows the recommend-
ed setup for this measurement. C1 and C2 block DC current flowing into the instrument. The value
of C1 should be calculated using the minimum measurement frequency. For example, when the
measurement is made at 1 kHz and above, C1 should be larger than 32 µF. The voltage rating of C1
and C2 must be higher than the output voltage of the battery.
Figure 5-37. Battery measurement setup
5-27