Figure 5-40. External DC bias measurement setup
External DC voltage bias protection in 4TP configuration
If the measurement frequency is above 2 MHz or the type of DUT is not suitable for these external
bias fixtures, it is recommended to use the protective circuit shown in Figure 5-41. This circuit is
usable with bias voltage up to ± 200 V. To reduce the effects of this additional circuit, perform the
open/short compensation with no bias voltage applied.
Figure 5-41. External DC voltage bias protection circuit
5-31