Agilent Technologies 4294A Specifications Page 99

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 98
5-10. Balanced device measurement
When a balanced DUT (such as balanced cable or the balanced input impedance of a differential
amplifier) is measured, it is necessary to connect a “balun” (balance-unbalance) transformer
between the instrument and the DUT. Looking from the DUT side, the UNKNOWN terminals of the
impedance measurement instrument are in an “unbalanced” configuration. Figure 5-35 (a) shows
an example of an auto balancing bridge instrument. Its Low terminal is considered a “virtual
ground” because it is held at approximately a 0 V potential. When a 1:1 balun transformer is con-
nected as shown in Figure 5-35 (b), the instrument can measure a balanced DUT directly.
An actual balun transformer has a limited frequency range. The measurement must be made within
its frequency range. In addition to Agilent’s balanced/unbalanced converters, various types of com-
mercial bulan transformers are available for various frequency ranges. To select the appropriate
balun transformer, check the frequency range and the impedance of the transformer’s balanced
(DUT) side. Its impedance should be close to the characteristic impedance of the DUT. The imped-
ance of the unbalanced side should be 50 or 75 as appropriate for the measurement instrument.
Open/short/load compensation for the balun transformer is required when the turns ratio of the
balun transformer used is not 1:1, or when an accurate measurement is needed. Open/short com-
pensation is not adequate because the balun transformer will produce both magnitude (|Z|) and
phase errors due to its transfer function characteristic. The terminal connectors of balanced side
should be connectable for both the standard devices used for open/short/load compensation and the
DUT. Figures 5-36 (a) through (d) show an example of an actual balun configuration and compensa-
tion.
Figure 5-35. Balanced device measurement
5-25
Page view 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 125 126

Comments to this Manuals

No comments