Agilent Technologies 4294A Specifications Page 37

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 36
SECTION 3
Fixturing and cabling
When interconnecting a device under test (DUT) to the measurement terminals of the auto balancing
bridge instrument, there are several connection configurations to choose from. This section will
introduce the basic theory and use of each connection configuration focussing on the auto balancing
bridge instrument. In RF impedance measurements, the usable connection configuration is the two
terminal method only. Since the measurement technique for RF impedance is different from that
for LF, it is described separately after the discussion of the auto balancing bridge instrument.
3-1. Terminal configuration
An auto balancing bridge instrument is generally equipped with four BNC UNKNOWN terminals (Hc,
Hp, Lp and Lc) on its front panel. There are several connection configurations used to interconnect
a DUT to the UNKNOWN terminals. Because each method has advantages and disadvantages, the
most suitable method must be selected based on the DUT’s impedance and required measurement
accuracy.
The two-terminal (2T) configuration is the simplest way but contains many error sources. Lead induc-
tance, lead resistance, and stray capacitance between two leads have been added to the measure-
ment result (Figure 3-1). Because of the existence of these error sources, the typical impedance
measurement range (without doing compensation) is limited to 100 to 10 k.
The three-terminal (3T) configuration employs coaxial cables to reduce the effects of stray capacitance.
The outer conductors (shield) of the coaxial cables are connected to the guard terminal.
Measurement accuracy is improved on the higher impedance measurement range but not on lower
impedance measurement range because lead inductance and resistance still remain. (See Figure 3-
2.) The typical impedance range will be extended above 10 k. If the outer conductor is connected
as shown in Figure 3-2 (d), lower impedance measurement accuracy is a little improved. (Shielded
2T configuration.)
The four-terminal (4T) configuration can reduce the effects of lead impedances because the signal cur-
rent path and the voltage sensing cables are independent (Figure 3-3). Accuracy for the lower
impedance measurement range is improved typically down to 1 . When the DUT’s impedance is
lower than 1 , a large signal current flows through the current path and mutual (M) coupling to the
voltage sensing cable will cause an error.
The five-terminal (5T) configuration is a combination of the 3T and 4T configurations. It is equipped
with four coaxial cables and all of the outer conductors of the four cables are connected to the guard
terminal (Figure 3-4). This configuration has a wide measurement range from 1 to 10 M, but the
mutual coupling problem still remains. If the outer conductor is connected as shown in Figure 3-4
(d), lower impedance measurement accuracy is a little improved. (Shielded 4T configuration.)
3-1
Page view 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 125 126

Comments to this Manuals

No comments