2. It is important to properly set the oscillator output level; resonators are test signal dependent.
The minimum impedance value and the series resonant frequency may vary depending on the
applied test signal level. Decrease the test signal level while monitoring the test current (I-mon-
itor function) until the specified test level is obtained.
3. Perform an open/short compensation. Use ALL POINT compensation mode instead of the inter-
polation mode because resonator measurements are narrowband. Also, pay special attention to
the short compensation procedure. Improper short compensation will result in measurement
error for fr and the minimum impedance value.
4. Keep the measurement temperature constant. Resonators are temperature sensitive. Place a
resonator into the test fixture with your hand and wait until the series resonant frequency
becomes constant.
Figure 5-31 (a) and (b) shows the measurement result for a crystal resonator.
Figure 5-29. Typical resonator characteristics
Figure 5-30. Resonator measurement setup