Agilent Technologies 4294A Specifications Page 44

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 43
3-4. Test cables
When the DUT is tested apart from the instrument, it is necessary to extend the test ports
(UNKNOWN terminals) using cables. If the cables are extended without regard to their length, it will
cause not only an error, but will also result in bridge unbalance making measurement impossible.
3-4-1. Agilent supplied test cables
Agilent Technologies supplies 1 m, 2 m and 4 m cables as listed in Table 3-2. The test cables from
16048A to 16048E are constructed using the same cable material. The 16048G and 16048H employ a
high quality cable to insure low-loss transmission characteristics that specifically match to the
4294A. The cable length and the usable frequency range must be considered when selecting a test
cable. Agilent’s instruments can minimize the effects of test cables because the error of the cable is
known. The measurement error will increase according to the cable length and the measurement
frequency.
Table 3-2. Agilent supplied test cables
3-4-2. User fabricated test cable
Using cables other than those supplied by Agilent is not recommended. The cable compensation
function of the instrument may not work properly in non-Agilent cables. If there is an unavoidable
need to use non-Agilent cables, then use the same or equivalent cable that is used to make the
Agilent test cables. The Agilent part number of the cable used for frequency below 30 MHz is 8120-
0367 (not applicable to the 4294A.) Electrical specifications for these cables are provided in Figure
3-8. Do not use test cables other than Agilent supplied cables for higher frequencies. To extend the
cables using the 4TP configuration, the cable length should be 1 m or 2 m so that the measurement
instrument can compensate for it (depending on the instrument’s cable length compensation func-
tion). If there is an error in the cable length, it will cause additional error. A detailed discussion is
provided in Section 4.
Test cable Cable Maximum Connector Applicable instruments
length frequency type
16048A 1 m 30 MHz BNC 4263B, 4268A, 4279A,
16048B SMC E4980A, 4284A, 4285A, 4288A
16048D 2 m 30 MHz BNC 4263B, 4268A, 4279A, E4980A, 4284A,
4285A, 4288A
16048E 4 m 2 MHz 4263B, E4980A, 4284A
16048G 1 m 110 MHz 4294A
16048H 2 m 110 MHz 4294A
3-8
Page view 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 125 126