Figure 5-33. Measurement result
Balanced cable measurement
A balun transformer is required for measuring balanced cable because the instrument’s UNKNOWN
terminal is unbalanced. Refer to the next paragraph titled “Balanced device measurement” for
details of measurement method. Figure 5-34 shows measurement setup for balanced cable. Agilent
16314A balanced/unbalanced 4-terminal converter can be used to measure balanced cables from
100 Hz to 10 MHz using an auto balancing bridge instrument. For measurement using a network
analyzer, the 16315A, 16316A and 16317A are available. These converters have different character-
istic impedance to allow impedance matching with DUT (cable) impedance of 50 Ω, 100 Ω and 600
Ω, respectively, as shown in Table 5-4.
Table 5-4. 16314A, 16315A, 16316A and 16317A characteristic impedance
Figure 5-34. Balanced cable measurement setup
5-24
Converter Characteristic impedance Applicable instrument
Unbalanced Balanced
side side
16314A 50 Ω 50 Ω Auto balancing bridge instrument
16315A 50 Ω 50 Ω Network analyzer
16316A 50 Ω 100 Ω
16317A 50 Ω 600 Ω