Agilent Technologies 4294A Specifications Page 35

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 34
2-7-3. Calibration
Most of the RF vector measurement instruments such as network analyzers need to be calibrated
each time a measurement is initiated or a frequency setting is changed. The RF I-V measurement
instrument requires calibration as well. At higher frequencies, a change in the instrument’s operat-
ing conditions, such as, environmental temperature, humidity, frequency setting, etc., have a greater
effect on measurement accuracy. This nature of RF vector measurement makes it difficult to suffi-
ciently maintain the calibrated measurement performance over a long period of time. Thus, users
have to periodically perform requisite calibration.
Note: Calibration is necessary each time a measurement setup is changed.
Calibration is executed in reference to three standard terminations, open, short and load, and all
three must be performed. To improve the accuracy of low dissipation factor measurements (high Q
factor), calibration with a low loss capacitor can be performed. The theory of calibration and appro-
priate calibration methods are discussed in Section 4.
2-7-4. Compensation
Two kinds of compensation functions are provided: open/short compensation for eliminating the
errors due to test fixture residuals and electrical length compensation for minimizing the test port
extension induced error. Practical compensation methods are discussed in Section 4.
2-7-5. Measurement range
RF I-V measurement method as well as network analysis covers the full measurement range from
low impedance to high impedance without ranging operation. All measurements are made at single
broad range.
2-7-6. DC bias
The internal DC bias source is connected to the center conductor of the coaxial test port and applies
a bias voltage to the DUT. The internal bias function can be set to either the voltage source mode or
the current source mode. The voltage source mode is adequate to the voltage-biased measurement
of capacitive DUT. The current source mode is to the current-biased measurement of inductive
DUT. Actual bias voltage and current across the DUT are monitored and, within specified
voltage/current output compliance ranges, automatically regulated at the same level as the bias set-
ting value regardless of the DUT’s DC resistance, thus allowing accurate DC bias to be applied
across the DUT. Since the internal bias source cannot output bias current large enough for inductor
measurements, generally, current-biased measurement (in excess of maximum output current)
requires an external bias method to be used. For biasing up to 5 A and 40 V in frequency range
below 1 GHz, the 16200B bias adapter compatible with RF I-V instruments is available.
2-21
Page view 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 125 126

Comments to this Manuals

No comments