Agilent Technologies 4294A Specifications Page 42

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 41
Table 3-1. Test fixture’s DUT connection configuration and applications
3-3-2. User fabricated test fixtures
If the DUT is not applicable to Agilent supplied test fixtures, create an application specific test fix-
ture. Key points to consider when fabricating a test fixture are:
(1) Residuals must be minimized. To minimize the residuals, the 4TP configuration should be main-
tained as close as possible to the DUT. Also, proper guarding techniques will eliminate the
effects of stray capacitance. See “Eliminating the stray capacitance effect” given in this section
for practical use of guarding.
(2) Contact resistance must be minimized. Contact resistance will cause additional error. In the case of
the 2T configuration, it directly affects the measurement result. The contact electrodes should
hold the DUT firmly and should always be clean. Use a corrosion-free material for the elec-
trodes.
(3) Contacts must be able to be opened and shorted. Open/short compensation can easily reduce the
effects of the test fixture residuals. To perform an open/short measurement, you must open
and short the contact electrodes. For an open measurement, the contact electrodes should be
located the same distance apart as when the DUT is connected. For the short measurement, a
lossless (low impedance) conductor should be connected between the electrodes, or contact
electrodes should be connected directly. If the four-terminal configuration is kept to the elec-
trodes, make the connections of current and potential terminals and then make an open or
short as shown in Figure 3-6.
3-6
DUT Applicable Agilent Basic characteristics Suitable
connection device type test fixture application
configuration
2-terminal Leaded device 16047D • Measurement is Middle and high
16047E susceptible to the effect impedance DUTs and high
16065A of residual impedance frequency measurements.
42842A/B/C and contact resistance
Surface mounted 16034E • Usable frequency limit
device 16034G is high
16034H • Additional error at
16334A high frequencies is
Material 16451B smaller than in
16452A 4-terminal connection
In-circuit device 16095A
42941A
4-terminal Leaded device 16047A • Measurement is less Low and middle
16089A/B/C/D/E affected by residual impedance DUTs and low
Surface mounted 16044A impedance and contact frequency measurements.
device resistance (at relatively
low frequencies)
• Usable frequency limit
is low
• Additional error at
high frequencies is
greater than in
2-terminal connection
Page view 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 125 126

Comments to this Manuals

No comments