92 3
Adjustment
HPOT-Amplifier Offset Correction Constants
H
POT
-Amplifier Offset Correction Constants
The purpose of this procedure is to adjust the H
POT
-amplifier offset.
Required Equipment
Description Recommended Model
4TP OPEN termination 42090A
Procedure
1. Run the adjustment program.
2. Choose the Adj_HpAmp_Ofst.
3. Connect the OPEN termination as shown in Figure 3-6
Figure 3-5 Relative Impedance Correction Constants Setup
4. Follow the adjustment program instruction to update the correction constants.