72 2
Performance Test
Performance Test Record
Measurement Accuracy Test
Accessory: Nothing
Standard: 1 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 MHz Cp ±0.0015 pF pF ±0.0005 pF
D ±0.0015 ±0.0002
3 MHz Cp ±0.0057 pF pF ±0.0010 pF
D ±0.0057 ±0.0008
Accessory: Nothing
Standard: 10 pF
OSC Level Frequency Parameter Test Limit Te st Re sult Measurement
Uncertainty
0.5 V 1 kHz Cp ±0.41 pF pF ±0.0010 pF
D ±0.041 ±0.0001
1 MHz Cp ±0.0083 pF pF ±0.0025 pF
D ±0.0008 ±0.0002
3 MHz Cp ±0.033 pF pF ±0.0030 pF
D ±0.0033 ±0.0005
10 MHz Cp ±0.101 pF pF ±0.015 pF
D ±0.0101 ±0.003
0.1 V 1 kHz Cp ±2.00 pF pF ±0.0010 pF
D ±0.200 ±0.0001