Performance Test
Performance Test
Figure 2-2 OSC Level Accuracy Test Setup 1
2. Press Reset key (blue,) to initialize the multimeter, then Set it as follows.
a. Press [ACV] to set the measurement mode to AC voltage.
b. Press S(blue - N Rdgs/Trig),,, to display SETACV.
c. Press ,,, to display SYNC, then press [Enter]
d. Press [NPLC] - [1] - [0] - [0] - [Enter]
3. Press [Preset] to initialize the 4294A. Then set the controls as follows.
Setting Operation
Frequency Span: 0 Hz [Span] - [0] - [×1]
Center Frequency: 1 kHz [Center] - [1] - [k/m]
OSC Level: 1 V [Source] - LEVEL - [1] - [×1]
OSC Level Monitor: ON [Display] - OSC MON
Number of Points [Sweep] - NUMBER OF POINTS - [2] - [×1]
4. Press [Trigger] - Single for a single sweep measurement:
5. Record the multimeter reading and the OSC Level Monitor Reading to the calculation
sheet.
6. Calculate the monitor accuracy according to the calculation sheet, and record the
monitor accuracy into the calculation sheet.
NOTE OPEN Reading is recorded in the Step22.