Relative Impedance Correction Constants
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
Relative Impedance Correction Constants
The purpose of this procedure is to adjust the relative impedance between the range
resistors.
Required Equipment
Description Recommended Model
4TP OPEN termination 42090A
Procedure
1. Run the adjustment program.
2. Choose the Adj_Relative_Z.
3. Connect the OPEN termination as shown in Figure 3-6
Figure 3-6 Relative Impedance Correction Constants Setup
4. Follow the adjustment program instruction to update the correction constants.