Agilent Technologies 4294A Service Manual Page 157

  • Download
  • Add to my manuals
  • Print
  • Page
    / 249
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 156
7 149
Service Related Menus
DIAGNOSTIC TEST Menu
7. Service Related Menus
Figure 7-1 Test Status on the Display
To see the test status of the desired test, enter the desired test number using the numeric
keypad, [], [], or RPG knob. Also, the GPIB commands listed below are available to get
the status using GPIB.
:DIAG:TEST:RES? <numeric> returns the test status. The <numeric> specifies
the test number and is an integer from 0 to 31.
*TST? executes internal test 0: ALL INTERNAL and
returns the test result.
Table 7-1 shows the test status abbreviation, its definition, and the GPIB test status code.
Table 7-1 Te s t Status Terms
Status Abbreviation Definition GPIB Code
PASS Pass “PASS”
FAIL Fail “FAIL”
BUSY In progress “BUSY”
NDON Not done “NDON”
DONE Done “DONE”
The test status is not stored in nonvolatile memory (battery backup memory). When the
analyzer is turned off and on, the analyzer sets test status abbreviation except the power-on
self test (internal tests 5 to 16) to “NDONE” (Not done). If a test is aborted by pressing any
key during its execution, the test status is undefined.
Test Descriptions
This section describes all 31 diagnostic tests.
Page view 156
1 2 ... 152 153 154 155 156 157 158 159 160 161 162 ... 248 249

Comments to this Manuals

No comments