Page view 118
4 111
Troubleshooting
Internal Tests Failure Troubleshooting
4. Troubleshooting
Internal Tests Failure Troubleshooting
Check the Power-On Selftest
The analyzer performs the power-on selftest every time the analyzer is turned on. In the
power-on selftest, internal tests 5 through 16 are executed sequentially. The first failed test
indicates the most probable faulty assembly and should be replaced. For more information
about the internal tests, see Chapter 7.
If the power-on selftest fails and the “POWER ON TEST FAILED” message is displayed,
execute all internal tests to identify the first failed test. Refer to the following steps to
execute all internal tests. Then refer to
Table 4-1 to know what assembly you should
replace.
Execute All Internal Tests
If the power-on self test fails and the “POWER ON TEST FAILED” message is displayed,
you should execute the all internal tests and identify the first failed test. To execute all
internal tests, you should specify test number as 0 in the HP
4294A diagnostic test mode.
1. Press [Preset], [System], DIAGNOSTIC TEST, [0], and [x1] to access the internal test
0: ALL INT.
2. Press RUN TEST to execute the ALL INT test.
3. Wait until the test result, PASS or FAIL, is displayed.
4. Press the [], [] keys to find the first occurrence of a FAIL message.
Execute the Individual Test
If you would like to perform the internal test individually, you should specify internal test
number in the HP
4294A diagnostic test mode.
1. Press [Preset], [System], DIAGNOSTIC TEST. Pressing INTERNAL TEST jumps to
the first test internal test, test:0 ALL INT.
2. Select the test using the RPG knob, [], [] keys or ENTRY keys.
3. Press RUN TEST to execute the specified test.
4. Wait until the test result, PASS or FAIL, is displayed.
NOTE Internal tests 1, 2, 3, and 4 are not executed in the power-on selftest. You should perform
these tests by specifying test number or by executing all internal tests.
NOTE The troubleshooting procedures requiring the internal tests 1,2, and 4 are also given in
Chapter 5.