228 Appendix C
Error messages
Order of Alphabet
138 DCBIAS CONSTANT OPERATION FAILED
When the DC bias mode was constant voltage (VOLT CONSTANT) or constant current
(CURRRENT CONSTANT), the actual DC bias output did not converge to the set DC
voltage value or DC current value, or it exceeded the set voltage limit or current limit. If a
DUT needs time to be stabilized after DC bias is applied, specify the point delay. If
necessary, change the voltage limit value or the current limit value.
139 DCBIAS LEVEL MONITOR NOT VALID
When the DC bias level monitor was not on, you attempted to execute the command to turn
on the marker DC voltage level monitor or the marker DC current level monitor
(MKRMON DCV or MKRMON DCI) and, as a result, the command was ignored. Before
turning on the marker DC voltage level monitor or the marker DC current level monitor,
turn on the corresponding DC bias level monitor (command: BMON VOLT or BMON
CURR).
137 DCBIAS OVERLOAD
When the DC bias was applied, DC current that exceeded the setting of the DC bias range
(<1mA, <10mA, or <100mA) was supplied from the bias source, or the overcurrent limit
circuit of the DC bias source operated. Increase the setting of the DC bias range, or
decrease the level of the applied DC bias.
99 DELTA MARKER OFF
When the delta marker was not displayed, you attempted to execute a function that needed
the display of the delta marker and, as a result, the command was ignored. For example,
when the delta marker was not displayed, a command corresponding to the MKRD?SPAN
key (MKRDSPAN) was sent. Before executing a function using the delta marker, first
display the delta marker (command: DMKR ON).
118 DEVICE TYPE IS NOT DOS
The execution of write access to the mass storage could not be executed because the format
type of the mass storage was not DOS.
37 DISPLAY BUFFER IS FULL
The use of the DRAW or MOVE command of HP Instrument BASIC made 4294As
display buffer full. Further write to the display buffer is impossible.
E
194 EEPROM WRITE FAIL
A write error was detected in the memory (EEPROM) to store the OPEN, SHORT, and
LOAD data for the four-terminal pair extension setting. Contact your local Agilent
Technologies sales office or the company you purchased this instrument from.
-200 Execution error
A comprehensive execution error occurred for which the 4294A could not detect further
details of the error. This error code simply indicates the occurrence of an execution error
that is defined in IEEE488.2,11.5.1.1.5.
-123 Exponent too large
The absolute value of the exponent exceeds 32,000. (Refer to IEEE488.2,7.7.2.4.1.)
-178 Expression data not allowed