Agilent Technologies 4294A Service Manual Page 114

  • Download
  • Add to my manuals
  • Print
  • Page
    / 249
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 113
106 4
Troubleshooting
TROUBLESHOOTING SUMMARY
TROUBLESHOOTING SUMMARY
The troubleshooting strategy of this manual is based on a verification (rather than
symptomatic) approach. This chapter describes typical troubleshooting procedure step by
step and you can identify a faulty assembly or a faulty group by following the
troubleshooting steps sequentially.
Figure 4-1 diagrams the troubleshooting organization.
Inspecting the power on sequence and the rear panel feature indicates a faulty group
(power supply or digital control). A faulty assembly is isolated to the faulty group accoding
to the troubleshooting procedures for each faulty group. The troubleshooting procedures
are given in Chapter 5 and 6.
Performing the internal tests, the external tests or the performance tests indicates a faulty
assembly. Assemblies of the analog section (source, transducer and vector ratio detector) of
the analyzer are verified by performing these tests.
A faulty assembly is replaced according to Chapter 9. Chapter 9 lists the replaceable parts
and gives replacement procedures for the major parts.
The procedures required after assembly replacement, such as adjustments and performance
tests, are given in Chapter 10.
Page view 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 248 249

Comments to this Manuals

No comments