Agilent Technologies 4294A Service Manual Page 85

  • Download
  • Add to my manuals
  • Print
  • Page
    / 249
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 84
2 77
Performance Test
Performance Test Record
2. Performance Test
Accessory: 42942A Terminal Adapter
Standard: 50 Ω
OSC Level Frequency Parameter Test L imi t Test Result Measurement
Uncertainty
0.5 V 1 kHz |Z| ±0.31 Ω Ω ±0.090 Ω
θ ±6.1 mrad mrad ±1.8 mrad
1 MHz |Z| ±0.31 Ω Ω ±0.090 Ω
θ ±6.1 mrad mrad ±1.8 mrad
10 MHz |Z| ±0.31 Ω Ω ±0.090 Ω
θ ±6.2 mrad mrad ±1.8 mrad
100 MHz |Z| ±0.54 Ω Ω ±0.010 Ω
θ ±10.9 mrad mrad ±2.0 mrad
Accessory: 42942A Terminal Adapter
Standard: 10 cm Airline with OPEN
OSC Level Frequency Parameter Test L imi t Test Result Measurement
Uncertainty
0.5 V 1 MHz |Z| ±155 Ω Ω ±44 Ω
θ ±7.1 mrad mrad ±2.0 mrad
10 MHz |Z| ±15.7 Ω Ω ±4.4 Ω
θ ±7.2 mrad mrad ±2.0 mrad
100 MHz |Z| ±2.32 Ω Ω ±0.43 Ω
θ ±10.8 mrad mrad ±2.0 mrad
Accessory: 42942A Terminal Adapter
Standard: 10 cm Airline with SHORT
OSC Level Frequency Parameter Test L imi t Test Result Measurement
Uncertainty
0.5 V 1 MHz |Z| ±0.0062 Ω Ω ±0.0017 Ω
θ ±54.4 mrad mrad ±15 mrad
10 MHz |Z| ±0.0165 Ω Ω ±0.0043 Ω
θ ±15.3 mrad mrad ±4.0 mrad
100 MHz |Z| ±0.157 Ω Ω ±0.027 Ω
θ ±14.7 mrad mrad ±2.5 mrad
Page view 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 248 249

Comments to this Manuals

No comments