2. Performance Test
Figure 2-9 Measurement Accuracy Test Setup (Accessory: None)
5. Press [Cal] - FIXTURE COMPEN - OPEN
6. Connect the SHORT termination instead of the OPEN termination
7. Press [Cal] - FIXTURE COMPEN - SHORT
8. Press [Meas] - more 1/3 - Cp-D to set the 4294A Measurement mode to Cp-D.
9. Create the sweep list shown in Figure 2-10. Segment 1 is added by the following
procedure.
a. Press [Sweep] - EDIT LIST - [ADD] to add a segment to the sweep list.
b. Edit the segment as follows.
Setting Operation
Start Frequency: 40 Hz START - [4] - [0] - [×1]
Stop Frequency: 1 kHz STOP - [1] - [k/m]
Number of Point: 2 NUMBER OF POINT - [2]
OSC level: 500 mV more 1/3 - OSC LEVEL - [5] - [0] - [0] - [k/m]
Bandwidth more 2/3 - BANDWIDTH - 5 PRECISE
NOTE Press the START or STOP Softkey. Do NOT press the front panel key printed “Start” or
“Stop”.