Performance Test
Performance Test
31. Repeat from Step 28 to 30 for all frequency settings in Table 2-3
Table 2-3 OSC Level Accuracy Test (Interface box) Settings 3
Frequency
1 MHz
10 MHz
15 MHz
16 MHz
50 MHz
110 MHz
32. Disconnect the test equipment from the 7mm connect, then connect the OPEN
termination to the 7mm connector.
33. Press [Source] - LEVEL - [.] - [5] - [×1] to set the 4294A OSC level to 0.5 V.
34. Press [Trigger] - Single for a single sweep measurement:
35. Read the level monitor reading for all OSC level and frequency setting in Table 2-3,
then record them to the calculation sheet.
36. Calculate the test results according to the calculation sheet, then record them to the test
record.