Agilent Technologies 4294A Service Manual Page 242

  • Download
  • Add to my manuals
  • Print
  • Page
    / 249
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 241
Appendix C 233
Error messages
Order of Alphabet
C. Error Messages
When the memory trace was not displayed, you attempted to set the function that needed
the display of the memory trace and, as a result, the command was ignored. For example, a
command (MKRO MEMO or SCAF MEMO) corresponding to the MARKER ON
[MEMORY] key display or the SCALE FOR [MEMO] key display was sent. Before
setting the function that needs the display of the memory trace, display the memory trace
(command: DISP
MEMO or DISP DATM).
-252 Missing media
You attempted to access the storage device, but media did not exit or was not installed
correctly. This message is displayed, for example, when you specifies the floppy disk drive
as the storage device but no floppy disk is set into the drive (correctly).
-109 Missing parameter
The number of parameters is less than that the command requires. For example, though the
“POIN” command requires 1 parameter like “POIN 101,” no parameter is added like
“POIN.”
36 MUST BE MORE THAN 2 POINTS FOR ANALYSIS
Though the calculation of equivalent circuit parameters (CALCULATE PARAMETERS
key, CALC PARMS key, or CALECPARA command) was executed, because the number of
points (NOP) within the sweep range (if the partial search function is on, within the
specified search range) is 2, they cannot be calculated. Set the number of points within the
sweep range (if the partial search function is on, within the specified search range) to 3 or
more.
N
98 NO ACTIVE MARKER
When the marker was not displayed, you attempted to execute the marker moving
(Marker?) function and, as a result, the command was ignored. For example, when the
marker was not displayed, a command corresponding to the MKR?START key
(MKRSTAR) was sent. Before executing the marker moving function, display the marker
(command: MKR ON).
0 (No error)
No error has occurred.
This message is not displayed on the LCD. 0 is returned as the error number if no error has
occurred in the instrument when the OUTPERRO? command is sent through GPIB.
100 NO FIXED DELTA MARKER
When the fixed Δ marker (FIXED DMKR) was not displayed, you attempted to execute a
command of a function that needed the display of the fixed Δ marker and, as a result, the
command was ignored. For example, when the fixed Δ marker was not displayed, a
command corresponding to the FIXED DMKR VALUE, FIXED DMKR AUX VALUEN
key (DMKRVAL <value>, DMRRAUV <value>) was sent. Before executing a function using
the fixed Δ marker, first display the fixed Δ marker (command: DMKR FIX).
113 NO STATE/DATA FILES ON DISK
Though you attempted to display the list of files and directories on a floppy disk on the
softkey label by executing the recall ([Recall]) key, file re-save (RE-SAVE FILE) key, file
deletion (PURGE FILE) key, directory change (CHANGE DIRECTORY) key, or file copy
Page view 241

Comments to this Manuals

No comments