Adjustment
DC Bias Level Correction Constants
DC Bias Level Correction Constants
The purpose of this procedure is to adjust the DC bias level.
Required Equipment
Description Recommended Model
Multimeter 3458A
Inter face Box p/n 04294-61002
100 Ω resistor
p/n 04294-61001
*1.An attachment to the 4294A
*1
BNC(m)-BNC(m) Cable, 61 cm p/n 8120-1839
Dual Banana-BNC(f) Adapter p/n 1251-2277
Procedure
1. Run the adjustment program.
2. Choose the Adj_DC_BIAS.
3. Connect the equipment termination as shown in Figure 3-10.
Figure 3-10 DC Bias Level Correction Constants Setup 1
4. Follow the adjustment program instruction to update the correction constants.