Performance Test
Performance Test
2. Performance Test
7. Confirm that the monitor accuracy is within the monitor accuracy limit.
NOTE If the monitor accuracy is beyond the limit, OSC Level Test fails.
8. Repeat from Step 4 to 7 for all OSC level and frequency setting in Table 2-1
Table 2-1 OSC Level Accuracy Test (Interface box) Settings 1
OSC Level Frequency
1.0 V 1 kHz
100 kHz
0.5 V 40 Hz
1 kHz
10 kHz
100 kHz
0.25 V 1 kHz
100 kHz
0.125 V 1 kHz
100 kHz
0.064 V 1 kHz
100 kHz
0.016 V 1 kHz
100 kHz
Accessory: Nothing, 1 MHz to 110 MHz
9. Connect the power sensor to the power meter. Calibrate the power meter for the power
sensor.
10. Set the power meter measurement mode to “Watt”
11. Disconnect the BNC(m)-BNC(m) Cable from the interface box, and connect the
equipment as shown in
Figure 2-3.