Adjustment
Order of Adjustment
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
Order of Adjustment
When performing more than one Adjustment, perform them in the order they appear in this
chapter. The procedures are presented in the following order.
Write ID
CAUTION Perform this adjustment only when the EEPROM is replaced. To perform it, select the
WRITE_ID on the adjustment program.
Initialize Correction Constant
CAUTION Perform this adjustment only when the EEPROM is replaced. To perform it, the Init_CC
on the adjsutment program.
Frequency Accuracy Correction Constants
Image Rejection Correction Constants
Lp-Mixer Correction Constants
Null Offset Correction Constants
Null Phase Correction Constants
Relative Impedance Correction Constants
DC Bias Monitor Correction Constants
DC Bias Level Correction Constants
OSC Level Monitor Correction Constants
OSC Level Correction Constants
Impedance Measurement Correction Constants
Frequecy Reference Adjustment