Agilent Technologies 4294A Service Manual Page 246

  • Download
  • Add to my manuals
  • Print
  • Page
    / 249
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 245
Appendix C 237
Error messages
Order of Alphabet
C. Error Messages
When the band width function was off, you attempted to execute a command to search the
cutoff point on the trace (WIDSIN or WIDSOUT) and, as a result, the command was
ignored. Before executing the cutoff point search of the band width function, turn on the
band width function (command: WIDT ON).
-330 Self-test failed
The result of the self-test was a failure. Contact your local Agilent Technologies sales
office or service center, or refer to the service manual.
-221 Setting conflict
A program data element complying with the syntax standard was analyzed but the 4294A
could not execute it at present.
-150 String data error
An error not included in error numbers between -151 and -159 occurred during the syntax
analysis of a string data element.
-158 String data not allowed
A string data element was received at where the 4294A did not accept any string data
elements.
-130 Suffix error
A suffix error.
An error other than -131 to -139 occurred in the notation of a suffix.
-138 Suffix not allowed
A suffix is added to a numeric value element to which no suffix can be added.
-134 Suffix too long
The suffix is too long.
The suffix contains the notation of 12 characters or more. (Refer to IEEE488.2,7.7.3.4.)
-102 Syntax error
There is a command or data type that cannot be recognized. For example, though the
4294A did not accept any strings, a string was received.
-310 System error
One of “system errors” defined for the 4294A occurred.
T
87 THETA MEASUREMENT NOT SELECTED
When the setting of the phase (q) was not active as a measurement parameter, you
attempted to execute a command to set the phase unit (PHAU DEG or PHAU RAD) or a
command to set the phase expansion display (EXPP ON or EXPP OFF) and, as a result, the
command was ignored. The active status of the phase (θ) setting means:
•|Z|-θ or |Y|-θ is selected as a measurement parameter and trace B is active.
COMPLEX Z-Y is selected as a measurement parameter, and the display format is
PLOAR (polar coordinates display format) for the selected active trace.