OSC Level Accuracy (with 42942A Terminal Adapter)
OSC
Level
Frequency Test Limits
Test Result Measurement
Uncertainty
TAR
0.5 V
1 MHz ± 51.3 mV mV ± 5.2 mV 19.6
10 MHz ± 53.5 mV mV ± 5.3 mV 20.0
15 MHz ± 54.8 mV mV ± 5.3 mV 20.5
16 MHz ± 55.0 mV mV ± 5.3 mV 20.6
50 MHz ± 63.5 mV mV ± 5.4 mV 21.0
110 MHz ± 78.5 mV mV ± 5.7 mV 19.0
Change 3
Change the Measurement Uncertainty value of DC Bias Level Accuracy Test (Page 70 to 71)
to the following information and appended TAR to this report.
DC Bias Monitor Accuracy Test
Bias
Level
Test Limits
Test Result Measurement
Uncertainty
TAR
0 V ± 5.0 mV mV ± 0.001 mV 4.0
25 V ± 55.0 mV mV ± 0.61 mV 53.4
40 V ± 85.0 mV mV ± 0.86 mV 89.8
-25 V ± 55.0 mV mV ± 0.61 mV 40.3
-40 V ± 85.0 mV mV ± 0.86 mV 58.7
0 mA ± 0.50 mV mV ± 0.001 mV 145.9
20 mA ± 0.75 mV mV ± 0.004 mV 107.2
100 mA ± 1.75 mV mV ± 0.016 mV 93.1
-20 mA ± 0.75 mV mV ± 0.004 mV 143.9
-100 mA ± 1.75 mV mV ± 0.016 mV 72.6
Bias
Level
Test Limits
Test Result Measurement
Uncertainty
TAR
0 V ± 5.0 mV mV ± 0.001 mV 96.4
25 V ± 30.3 mV mV ± 0.61 mV 49.2
40 V ± 45.6 mV mV ± 0.86 mV 53.0
-25 V ± 30.3 mV mV ± 0.61 mV 49.2
-40 V ± 45.6 mV mV ± 0.86 mV 53.0
0 mA ± 0.20 mV mV ± 0.001 mV 43.8
20 mA ± 0.60 mV mV ± 0.004 mV 90.5
100 mA ± 2.20 mV mV ± 0.016 mV 115.3
-20 mA ± 0.60 mV mV ± 0.004 mV 73.4
-100 mA ± 2.20 mV mV ± 0.016 mV 93.2
Change 4
Change the Measurement Uncertainty value of Measurement Accuracy Test (Page 72 to 80)
to the following information and appended TAR to this report.
Measurement Accuracy Test