Theory of Operation
DIGITAL CONTROL OPERATION
terminal, the PRINTER port, the 24-BIT I/O PORT, and the LAN port.
A51 GSP
The A51 GSP (graphics system processor) provides an interface between the A1 CPU and
the A52 LCD. The A1 CPU converts the formatted data to GSP commands and writes them
to the A51 GSP. The A51 GSP processes the data to obtain the necessary signals and sends
these signals to the A52 LCD.
The A51 GSP receives two power supply voltages: + 5VD, which is used for data
processing and converted to + 3.3 V, and + 15 V, which is passed on the A54 Inverter. The
+ 3.3 V goes to the A52 LCD.
The A51 also has a 24 bit I/O control circuit, which interfaces the 4294A to an external
handler through the A34 I/O connector.
A54 Inverter
The A54 Inverter is located in the LCD module on the front panel assembly. The A54
receives + 15 V from the A51 GSP and provides a high voltage (800 to 1000 VAC) to the
backlight of the LCD.
A52 LCD (Liquid Crystal Display)
The A52 LCD is a 8.5 TFT Color LCD, receives a high voltage (800 to 1000 VAC) from
the A54 Inverter as backlight power and the digital horizontal and the vertical signals from
the A51 GSP.
A53 FDD
The 4294A has a built-in, 3-1/2 inch FDD (Flexible Disk Drive) on the front panel. The
high density or 2 double density 3-1/2 inch flexible disk is applicable. The A53 FDD reads
and writes data to and from the flexible disk.